Quantification of trunnion damage in a series of intact total hip arthroplasty femoral stems previously identified to be at risk of catastrophic failure

Brent A Lanting,
Olawale A Sogbein,
Steven J MacDonald
et al.

Abstract: Background: Corrosion at the head-neck junction of femoral stems is a rare complication of total hip arthroplasty (THA) with manifestations ranging from subclinical wear to failure. Prior studies have identified a single femoral component design with an increased propensity for catastrophic trunnion failure. The purpose of the present study was to quantify trunnion damage of this femoral component retrieved from patients undergoing revision THA for non-trunnionosis indications. Methods: 24 femoral components f… Show more