Quality assurance tests for the Gamma Knife® Icon™ image guidance system

Abstract: IntroductionThe Gamma Knife® Icon™ comes with an image guidance system for tracking patient motion and correcting for inter‐ and intrafractional shifts, mainly used with frameless thermoplastic immobilization. The system consists of a cone‐beam CT (CBCT) and a couch‐mounted infrared camera (IFMM). We report our quality assurance program for Icon's image guidance system.MethodsThe manufacturer‐provided tool is used for daily checks of CBCT positional precision. Catphan® phantom is used for monthly checks of CBC… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

3
20
0

Year Published

2019
2019
2022
2022

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 14 publications
(23 citation statements)
references
References 7 publications
(10 reference statements)
3
20
0
Order By: Relevance
“…The maximum deviations observed for the four BBs are 0.13, 0.12, 0.15, and 0.12 mm, respectively. These numbers are consistent with a previous study in which the maximum deviations from daily CBCT precision tests for two 2‐yr period were reported …”
Section: Resultssupporting
confidence: 93%
See 4 more Smart Citations
“…The maximum deviations observed for the four BBs are 0.13, 0.12, 0.15, and 0.12 mm, respectively. These numbers are consistent with a previous study in which the maximum deviations from daily CBCT precision tests for two 2‐yr period were reported …”
Section: Resultssupporting
confidence: 93%
“…These numbers are consistent with a previous study in which the maximum deviations from daily CBCT precision tests for two 2-yr period were reported. 5 Table I gives the X, Y, Z coordinates (referred to the axes of the Gamma Knife patient positioning system) 23 and the pitch, yaw, roll angles from the image registrations for manual shifts of the STEEV head phantom in the Z direction up to 10 mm. The result for 0 mm shift was obtained when the phantom was in the same position as the reference scan.…”
Section: C Effect Of Fractionationmentioning
confidence: 99%
See 3 more Smart Citations