2013
DOI: 10.1118/1.4817481
|View full text |Cite
|
Sign up to set email alerts
|

Quality assurance of proton beams using a multilayer ionization chamber system

Abstract: Our investigation shows that the Zebra can be useful not only for fast but also for accurate measurements of the depth-dose distributions of both scattered and scanned proton beams. The analysis of a large set of measurements shows that the commonly assessed beam quality parameters obtained with the Zebra are within the acceptable variations specified by the manufacturer for our delivery system.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

3
51
0

Year Published

2014
2014
2021
2021

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 48 publications
(54 citation statements)
references
References 26 publications
(19 reference statements)
3
51
0
Order By: Relevance
“…The Zebra device (IBA Dosimetry) is a commercial MLIC designed to measure depth‐dose curves under broad field conditions (10) . It consists of a stack of 180 independent air‐vented, plane‐parallel ionization chambers fabricated with printed circuit board (PCB) technologies (see Fig.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…The Zebra device (IBA Dosimetry) is a commercial MLIC designed to measure depth‐dose curves under broad field conditions (10) . It consists of a stack of 180 independent air‐vented, plane‐parallel ionization chambers fabricated with printed circuit board (PCB) technologies (see Fig.…”
Section: Methodsmentioning
confidence: 99%
“…Note that a 10×10cm2 reference field also matches the optimum irradiation conditions of the Zebra (10) . Fields are delivered in a standalone mode, as described for the Giraffe measurements above.…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…Range and SOBP measurements can be accurately done in a single measurement with a stack of scintillating plates, such as the 28×3.15emmm thick plastic scintillators presented by Amaldi et al (1) or using a Zebra (IBA Dosimetry, Schwarzenbruck, Germany), a commercial multilayer ionization chamber (2) . It is also possible to measure the output with a single measurement using an ionization chamber in a solid phantom, or the collinearity of the proton beam axis and X‐ray beam axis using a film or ionization chamber array.…”
Section: Introductionmentioning
confidence: 99%
“…The central-axis depth dose distribution in case (c) was measured by a multi-layer ionization chamber (MLIC, IBA Dosimetry GmbH, Germany) [22]. The 2D dose distributions at specific depths were measured with a 2D ion chamber array (MatriXX PT, IBA Dosimetry GmbH, Germany) [23].…”
Section: Measurementsmentioning