Quality assurance for image‐guided radiation therapy utilizing CT‐based technologies: A report of the AAPM TG‐179

Abstract: A program for quality control of CT-based image-guidance systems has been produced, with focus on geometry, image quality, image dose, system operation, and safety. Agreement and clarification with respect to reports from the AAPM TG-101, TG-104, TG-142, and TG-148 has been addressed.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

4
234
0
8

Year Published

2015
2015
2020
2020

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 271 publications
(267 citation statements)
references
References 201 publications
4
234
0
8
Order By: Relevance
“…The same is placed at the physical isocenter. A specific displacement of the couch from isocenter using a random number generator is applied (the movement in all axis is within 2 cm to be in agreement with TG‐179 (2) ). Then, a localization CBCT image dataset to assess the couch motion required to align the phantom with a reference CT dataset is acquired.…”
Section: Methodsmentioning
confidence: 96%
See 3 more Smart Citations
“…The same is placed at the physical isocenter. A specific displacement of the couch from isocenter using a random number generator is applied (the movement in all axis is within 2 cm to be in agreement with TG‐179 (2) ). Then, a localization CBCT image dataset to assess the couch motion required to align the phantom with a reference CT dataset is acquired.…”
Section: Methodsmentioning
confidence: 96%
“…

Test 1: Imaging and treatment coordinate coincidence. After the initial geometric calibration has been performed using procedures similar to that described by Bissonnette et al, (2) a daily geometric accuracy QA consists of stability and consistency tests is performed. The MIMI phantom (Standard Imaging Inc., Middleton WI), which has embedded BBs, was used to take portal images and CBCT, and the positions of the embedded BBs on these daily images as compared to the base line were used to verify the imager's consistency.

Test 2: Couch position and repositioning.

…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…Recommendations for IGRT system QA are provided in AAPM TG‐179 report 4. This includes the recommendation for testing the imaging and treatment isocenters coincidence.…”
Section: Introductionmentioning
confidence: 99%