Quality assessment of two lactate test strip methods suitable for obstetric use

Abstract: Accuracy of lactate determinations in cord blood was tested for one reflectometric (Accusport) and one amperometric (Lactate Pro) microvolume test strip lactate meter. Both meters, using a whole blood sample, measure lower levels of lactate than a reflectometric device considered as a reference method, which analyses lactate in plasma. Readings were unaffected irrespective of lactate concentrations for the Lactate Pro, whilst the Accusport overestimated low lactate concentrations and underestimated high values… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

2
34
0
3

Year Published

2001
2001
2017
2017

Publication Types

Select...
5
2
1

Relationship

0
8

Authors

Journals

citations
Cited by 44 publications
(39 citation statements)
references
References 11 publications
(14 reference statements)
2
34
0
3
Order By: Relevance
“…18,19 Studies that have used BlandAltman analysis showed that mean differences (95% LOA) were similar to our findings, Ϫ0.2 (Ϫ1.2 to 0.8), 16 Ϫ0.3 (Ϫ1.33 to 0.69), 18 and for a study that tested 2 Accusport machines with machine A Ϫ0.4 (Ϫ1.8 to 1.1) and machine B Ϫ0.4 (Ϫ1.5 to 0.7). 19 These published limits of agreement were also similar to the values for our Accusport B, although different for those of Accusport A.…”
Section: Discussionsupporting
confidence: 89%
See 1 more Smart Citation
“…18,19 Studies that have used BlandAltman analysis showed that mean differences (95% LOA) were similar to our findings, Ϫ0.2 (Ϫ1.2 to 0.8), 16 Ϫ0.3 (Ϫ1.33 to 0.69), 18 and for a study that tested 2 Accusport machines with machine A Ϫ0.4 (Ϫ1.8 to 1.1) and machine B Ϫ0.4 (Ϫ1.5 to 0.7). 19 These published limits of agreement were also similar to the values for our Accusport B, although different for those of Accusport A.…”
Section: Discussionsupporting
confidence: 89%
“…A number of other studies have examined the accuracy of the Accusport lactate analyzer in athletes, 14,15 neonates, 16,17 and critically ill patients. 18,19 Studies that have used BlandAltman analysis showed that mean differences (95% LOA) were similar to our findings, Ϫ0.2 (Ϫ1.2 to 0.8), 16 Ϫ0.3 (Ϫ1.33 to 0.69), 18 and for a study that tested 2 Accusport machines with machine A Ϫ0.4 (Ϫ1.8 to 1.1) and machine B Ϫ0.4 (Ϫ1.5 to 0.7).…”
Section: Discussionmentioning
confidence: 99%
“…Limitations include the use of fetal umbilical cord blood rather than scalp blood to allow sufficient volume, as others have; 16,19,20 studies using both have shown greater variation in scalp than cord samples, possibly due to mixture with tissue fluid. 13,26 Only two alternative POC meters which have undergone recent obstetric investigation 13,17,24 were included in the study, although other lactate devices are currently being studied for obstetric use, including Roche Accutrend PlusTM, 21 together with a range of devices in the sporting world.…”
Section: F I G U R Ementioning
confidence: 99%
“…These micromethods are simple, cheap and suitable for use in the delivery room [5] . Nordström et al [7] reported several factors that may alter the reliability of lactate determination. The reflectometric method gives falsely low levels, owing to slower diffusion of plasma in samples with a high hematocrit [7] .…”
mentioning
confidence: 99%
“…Nordström et al [7] reported several factors that may alter the reliability of lactate determination. The reflectometric method gives falsely low levels, owing to slower diffusion of plasma in samples with a high hematocrit [7] . Indeed, lactate concentrations are higher in plasma than in whole blood containing intact erythrocytes [1] .…”
mentioning
confidence: 99%