1987
DOI: 10.1001/archotol.1987.01860030090016
|View full text |Cite
|
Sign up to set email alerts
|

Proximal Large-Bore Jet Ventilation for Laryngeal Laser Surgery

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
6
0

Year Published

1987
1987
2001
2001

Publication Types

Select...
5
2

Relationship

0
7

Authors

Journals

citations
Cited by 26 publications
(6 citation statements)
references
References 21 publications
0
6
0
Order By: Relevance
“…Delivered by this method, into the open proximal end of 5-mm x 30-cm and 8mm X 40-cm bronchoscope s tripled and quadrupled the tidal volume, respectively, compared with the injector alone. This is due to entrainment of room air ahead of the jetted gas 66 (Fig. 4).…”
Section: Jet Ventilation Anesthesiamentioning
confidence: 99%
See 2 more Smart Citations
“…Delivered by this method, into the open proximal end of 5-mm x 30-cm and 8mm X 40-cm bronchoscope s tripled and quadrupled the tidal volume, respectively, compared with the injector alone. This is due to entrainment of room air ahead of the jetted gas 66 (Fig. 4).…”
Section: Jet Ventilation Anesthesiamentioning
confidence: 99%
“…Gastric distension is a common complication of supraglottic jetting, due to overflow of gas into the esophagus. 65 Maintaining a low insufflation pressure, i.e., 20 psi vs. 50 psi, was advised by Koufman, et al 66 Subglottic jet ventilation. Subglottic jet ventilation was first performed with 3.5-mm OD plastic chest tubes or by inserting the metal supraglottic tube through the glottis below the level of the vocal cords.0 7 -10 The end of a plastic jet tube tends to whip around in the trachea.…”
Section: Supraglottic Jet Ventilation the Originalmentioning
confidence: 99%
See 1 more Smart Citation
“…1). 1 Adequate exposure is obtained using an adult‐sized Jako laryngoscope (catalogue No. 52‐2200, Pilling Co., Fort Washington, PA).…”
Section: Methodsmentioning
confidence: 99%
“…Advantages of jet ventilation are similar to those ofspontaneous respiration. Negative aspects, however, include the need for constant alignment ofthe laryngoscope, larynx, and trachea; the risk of barotrauma (pneumothorax or pneumomediastinum) or gastric distention and aspiration; and the distal seeding of resected materials (17). As with spontaneous respiration, the trachea and vocal cords are not protected from stray beams (3).…”
Section: Airway Managementmentioning
confidence: 99%