2015
DOI: 10.1209/0295-5075/109/18005
|View full text |Cite
|
Sign up to set email alerts
|

Protein-based nanobiosensor for direct detection of hydrogen sulfide

Abstract: The chemically modified cytochrome c from equine heart, EC (232-700-9), was immobilized onto gold nanoparticles in order to develop a specific biosensing system for monitoring hydrogen sulfide down to the micromolar level, by means of a localized surface plasmon resonance spectroscopy. The sensing mechanism is based on the cytochrome-c conformational changes in the presence of H2S which alter the dielectric properties of the gold nanoparticles and the surface plasmon resonance peak undergoes a redshift. Accord… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
6
0

Year Published

2016
2016
2020
2020

Publication Types

Select...
6

Relationship

4
2

Authors

Journals

citations
Cited by 12 publications
(6 citation statements)
references
References 40 publications
(31 reference statements)
0
6
0
Order By: Relevance
“…LOD is defined as the lowest possible concentration of an analyte that can be distinguished without any assurance about the imprecision of the obtained results. Herein, the LOD was calculated by the following equation : LOD =3σ/bwhere b is the sensitivity of the sensor (slope of the regression line) and σ is the SD of the blank. Mean value and SD of Δip were calculated, since each concentration measurement was repeated three times (Figure C and D).…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…LOD is defined as the lowest possible concentration of an analyte that can be distinguished without any assurance about the imprecision of the obtained results. Herein, the LOD was calculated by the following equation : LOD =3σ/bwhere b is the sensitivity of the sensor (slope of the regression line) and σ is the SD of the blank. Mean value and SD of Δip were calculated, since each concentration measurement was repeated three times (Figure C and D).…”
Section: Resultsmentioning
confidence: 99%
“…LOD is defined as the lowest possible concentration of an analyte that can be distinguished without any assurance about the imprecision of the obtained results. Herein, the LOD was calculated by the following equation [37]:…”
Section: Detection Of the Total And Free Psamentioning
confidence: 99%
“…The k L can be described as follow : knormalL=k off +k on false[ DOX false]where k off is the rate of dissociation of the DOX from the GO–DOX complex and k on is the second‐order rate constant for the association of DOX and GO; the extracted values of Fig. are presented in Table 4.…”
Section: Resultsmentioning
confidence: 99%
“…The PLGA MPs size was determined using an image analysis method as described previously. 13 Five different microscopy images were prepared for each sample, and the average diameters and size distributions were determined by ImageJ soware (ImageJ freeware, NIH, USA). The hydrodynamic diameter size distribution of CQDs sample was measured by Dynamic Light Scattering (DLS, Malvern ZEN 3600 zetasizer) analysis.…”
Section: Vegf Bioactivity Assaymentioning