Prostate SBRT using O‐Ring Halcyon Linac — Plan quality, delivery efficiency, and accuracy

Abstract: Cone beam CT‐guided prostate stereotactic body radiotherapy (SBRT) treatment on the recently installed novel O‐ring coplanar geometry Halcyon Linac with a single energy 6MV‐flattening filter free (FFF) beam and volumetric modulated arc therapy (VMAT) is a fast, safe, and feasible treatment modality for early stage low‐ and intermediate‐risk prostate cancer patients. Following the RTOG‐0938 compliance criteria and utilizing two‐full arc geometry, VMAT prostate SBRT plans were generated for ten consecutive patie… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
5

Citation Types

2
29
0

Year Published

2021
2021
2023
2023

Publication Types

Select...
5

Relationship

1
4

Authors

Journals

citations
Cited by 16 publications
(32 citation statements)
references
References 40 publications
2
29
0
Order By: Relevance
“…Most importantly, these statistically significant differences were not observed when benchmarking HA2‐DL10 and HA3‐DL10 with the C‐arm linacs as all these planning techniques demonstrated comparable plan quality (Figure 3). This latter observation is consistent with the results obtained by Pokhrel et al, which compared the plan quality of dual‐arc VMAT on the fast‐rotating O‐ring linac with dual‐arc VMAT on a conventional C‐arm linac for whole‐gland prostate SBRT and also found no statistically significant differences 44 …”
Section: Discussionsupporting
confidence: 92%
See 3 more Smart Citations
“…Most importantly, these statistically significant differences were not observed when benchmarking HA2‐DL10 and HA3‐DL10 with the C‐arm linacs as all these planning techniques demonstrated comparable plan quality (Figure 3). This latter observation is consistent with the results obtained by Pokhrel et al, which compared the plan quality of dual‐arc VMAT on the fast‐rotating O‐ring linac with dual‐arc VMAT on a conventional C‐arm linac for whole‐gland prostate SBRT and also found no statistically significant differences 44 …”
Section: Discussionsupporting
confidence: 92%
“…All treatment plans were found to be within the tolerance limits of the AAPM TG‐218 report for IMRT measurement‐based verification QA as portal image dosimetry measurements had γ 2%/2mm agreement scores above 95% (Table 5). 43 These patient‐ specific QA results are consistent with previous reports on the treatment delivery quality of VMAT on the fast‐rotating O‐ring linac 44,46,47 …”
Section: Discussionsupporting
confidence: 86%
See 2 more Smart Citations
“…Portal dosimetry gamma pass rates were within tolerance levels for the 2%/2 mm, 2%/1 mm, and 1%/1 mm criterion for all Halcyon plans. Mean ArcCHECK gamma pass rates were in the range of literature reported values with Pokhrel et al reporting mean 98.6 ± 1.5% and 98.3 ± 2.0% for the 2%/2 mm criterion in prostate stereotactic body radiotherapy for Halcyon VMAT and TrueBeam VMAT, respectively 22 …”