Prospective randomized comparative study between venturi and peristaltic pumps in WhiteStar Signature<sup>&reg;</sup> phacoemulsification machine

Abstract: PurposeTo compare intraoperative parameters between venturi and peristaltic pump in WhiteStar Signature® phacoemulsification machine using the bevel-down technique.SettingHospital Oftalmológico de Brasília, Brasília, Federal District, Brazil.DesignProspective, comparative, patient-masked study.MethodsThree hundred eyes were randomly assigned to have a phacoemulsification procedure with WhiteStar Signature® using either peristaltic (n=150 eyes) or venturi (n=150 eyes) pumps by a single surgeon (WTH). Elliptical… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
7
0

Year Published

2019
2019
2024
2024

Publication Types

Select...
5

Relationship

1
4

Authors

Journals

citations
Cited by 7 publications
(8 citation statements)
references
References 11 publications
0
7
0
Order By: Relevance
“…The two main types of pumps used in devices for cataract surgery are the Venturi pump and the peristaltic pump. A few available comparative studies showed that the Venturi pump requires smaller energy levels and a shorter time to complete cataract surgery than the peristaltic pump [ 38 , 39 ]. The Venturi type is also associated with a lower incidence of thermal corneal burns related to excessive thermal energy levels during the surgery [ 39 ].…”
Section: Discussionmentioning
confidence: 99%
“…The two main types of pumps used in devices for cataract surgery are the Venturi pump and the peristaltic pump. A few available comparative studies showed that the Venturi pump requires smaller energy levels and a shorter time to complete cataract surgery than the peristaltic pump [ 38 , 39 ]. The Venturi type is also associated with a lower incidence of thermal corneal burns related to excessive thermal energy levels during the surgery [ 39 ].…”
Section: Discussionmentioning
confidence: 99%
“…The continuous improvement of phacoemulsification fluidics and machines, handpiece tips, the addition of femtosecond laser to cataract procedures and its continuous evolution and the use of optimized constants for optical biometry have provided surgeons with tools to enhance their capabilities and to be granted with postoperative emmetropia. [1][2][3][4][5][6][7][8][9] Multifocal intraocular lenses were designed since the late 1980s but were well accepted since the last decade to provide adequate vision for far and near. Those first universally accepted intraocular lenses had a higher addition and were bifocal.…”
Section: Introductionmentioning
confidence: 99%
“…Studies suggest that peristaltic fluidics have better control and holdability. 6 Also, studies show that Venturi pumps are more efficient, use less total ultrasound, decrease case time, and decrease fluid usage. 6 The fluidic properties of the dual-durometer tubing were designed to facilitate improved chamber stability and to reduce surge, which allows surgeons to have improved holdability and followability with both pump systems.…”
Section: Discussionmentioning
confidence: 99%
“… 6 Also, studies show that Venturi pumps are more efficient, use less total ultrasound, decrease case time, and decrease fluid usage. 6 The fluidic properties of the dual-durometer tubing were designed to facilitate improved chamber stability and to reduce surge, which allows surgeons to have improved holdability and followability with both pump systems. In this study, the surgeons frequently switched between modes during various stages of the surgery, and the ability to do so may account for the high satisfaction scores with the VERITAS Vision System.…”
Section: Discussionmentioning
confidence: 99%