“…The method's accuracy was validated using the following certifi ed reference materials (CRM): Tea Leaves (INCT-TL-1), Mixed Polish Herbs (INCT-MPH-2) [Dybczyński et al, 2004] and Tea (NCS ZC 73014, China National Analysis Center for Iron and Steel, Beijing, China).…”
Section: Samples and Sample Preparationmentioning
confidence: 99%
“…The method's accuracy was validated using the following certifi ed reference materials (CRM): Tea Leaves (INCT-TL-1), Mixed Polish Herbs (INCT-MPH-2) [Dybczyński et al, 2004] and Tea (NCS ZC 73014, China National Analysis Center for Iron and Steel, Beijing, China).…”
Section: Samples and Sample Preparationmentioning
confidence: 99%
“…Data for samples were considered only if it matched with the certified values for the RMs. Mean±SD data for INCT MPH-2 along with certified/ information values [34], error percentage, relative standard deviation (RSD) percentage, and Z scores are listed in Table 3. It is observed that our data match well within ±5-10% of the certified values with the exception of a few elements such as Ce, Cl, Co, Cu, Hf, Hg, and Zn, but all Z score values are between ±1.90 except for Ce and Sm suggesting our data to be within 99% confidence limit.…”
Section: Resultsmentioning
confidence: 99%
“…These were stored in precleaned polyethylene capped bottles. RMs Peach leaves (SRM 1547) [33] and Mixed Polish Herbs (MPH-2) [34] procured from the National Institute of Standards and Technology, USA, and the Institute of Nuclear Chemistry and Technology (INCT), Poland, respectively, were used as such.…”
Section: Sample Collection and Preparationmentioning
confidence: 99%
“…In our laboratory the certification of the candidate reference materials is performed on the basis of world-wide interlaboratory comparison and proven data evaluation methodology [16,17,18]. The uncertainty associated with the central value combines the uncertainty due to dispersion of analytical results obtained by various methods (after rejection of outliers), i.e.…”