Prediction of low-addition segmented refractive intraocular lens position and deviation using anterior-segment optical coherence tomography
Norihiro Mita,
Mai Yamazaki,
Yusuke Seki
et al.
Abstract:This study aimed to develop and analyze the accuracy of predictive formulae for postoperative anterior chamber depth, tilt, and decentration of low-added-segment refractive intraocular lenses. This single-center, retrospective, observational study included the right eyes of 96 patients (mean age: 72.43 ± 6.58 years), who underwent a cataract surgery with implantation of a low-added segmented refractive intraocular lens at the Medical University Hospital between July 2019 and January 2021, and were followed up … Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.