Preclinical Evaluation of a Novel Sirolimus-Eluting Iron Bioresorbable Coronary Scaffold in Porcine Coronary Artery at 6 Months

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
33
1

Year Published

2019
2019
2022
2022

Publication Types

Select...
8

Relationship

2
6

Authors

Journals

citations
Cited by 34 publications
(34 citation statements)
references
References 28 publications
0
33
1
Order By: Relevance
“…In total, 48 swines that received IBS were assessed by QCA, OCT, and histology, and their results were compared to the same number of subjects receiving Xience (Abbott Vascular, Santa Clara, CA, USA) at 7 days to 6 months follow-up. IBS results showed no significant difference in area of stenosis compared to Xience up to 6 months, while scaffold or stent thrombosis was observed during the study [45].…”
Section: Sirolimus-eluting Iron Bioresorbable Coronary Scaffold (Ibs)mentioning
confidence: 75%
See 2 more Smart Citations
“…In total, 48 swines that received IBS were assessed by QCA, OCT, and histology, and their results were compared to the same number of subjects receiving Xience (Abbott Vascular, Santa Clara, CA, USA) at 7 days to 6 months follow-up. IBS results showed no significant difference in area of stenosis compared to Xience up to 6 months, while scaffold or stent thrombosis was observed during the study [45].…”
Section: Sirolimus-eluting Iron Bioresorbable Coronary Scaffold (Ibs)mentioning
confidence: 75%
“…Pure iron devices are highly biocompatible and structurally sturdy, however, they present a long corrosion period and slow clearance from the vessel up to 18 months from implantation [45]. However, iron-alloy processed with nitriding technology, a thermochemical heat treatment with nitrogenation of the iron piece, adds more reliability to corrosion rate when associated with PLLA coating as seen in porcine models at 13 months [45,46].…”
Section: Iron Alloy Platform Degradationmentioning
confidence: 99%
See 1 more Smart Citation
“…S1 ). It has comparable low crossing-profile, 5 French guiding catheter compatibility, mechanical strength, specification range, endothelialization, safety and effectiveness to state-of-the-art permanent drug-eluting stents [ 36 , 48 ] in vitro and in animal models. As presented in Fig.…”
Section: Discussionmentioning
confidence: 99%
“…No ScT was observed within 180 days after implantation. Tissue coverage of the IBS was more complete compared with that of the CoCr-EES at the 14 day follow-up ( 25 ). Interestingly, the IBS shows good device performance comparable to a current mainstream DES.…”
Section: Current Status Of Brssmentioning
confidence: 99%