Precision, accuracy, cross reactivity and comparability of serum indices measurement on Abbott Architect c8000, Beckman Coulter AU5800 and Roche Cobas 6000 c501 clinical chemistry analyzers

Abstract: There is heterogeneity between analyzers in the hemolysis, icteria, lipemia (HIL) quality performance. Verification of serum indices in routine work is necessary to establish analytical specifications.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
13
0

Year Published

2018
2018
2021
2021

Publication Types

Select...
9

Relationship

0
9

Authors

Journals

citations
Cited by 41 publications
(14 citation statements)
references
References 16 publications
0
13
0
Order By: Relevance
“…The Architect measures hemolysis at 500, 524, 572, 604, 628, 660, and 804 nm and returns a numerical value that approximately corresponds to g/L of free hemoglobin. The Architect hemolysis readings have proved to exhibit acceptable accuracy and reproducibility . Also, the used Architect instrument participates in an external quality program for HIL from EQUALIS AB .…”
Section: Methodsmentioning
confidence: 99%
“…The Architect measures hemolysis at 500, 524, 572, 604, 628, 660, and 804 nm and returns a numerical value that approximately corresponds to g/L of free hemoglobin. The Architect hemolysis readings have proved to exhibit acceptable accuracy and reproducibility . Also, the used Architect instrument participates in an external quality program for HIL from EQUALIS AB .…”
Section: Methodsmentioning
confidence: 99%
“…Generally, the normal hemoglobin concentration in human blood is 130-160 g/L for males and 120-150 g/L for females [30]; thus, the total IgE in completely hemolyzed blood samples of a normal human will be decreased by approximate 20% when analyzed through ECLIA due to interference, according to our results. A series of evaluation methods and indicators reflect the interference on different biomarkers and test methods in the clinical laboratories, such as the hemolytic index (HI), ± 10% change of concentrations from baseline (± 10%Δ) [14,31,32]. However, for postmortem biochemical analyses, more serious interfering factors are observed relative to clinical situations.…”
Section: Discussionmentioning
confidence: 99%
“…Samples were processed in Alinity c chemistry analysers (Abbott Laboratories, Illinois, USA and its quantitative results were given in g/L of haemoglobin (H-index), µmol/L of bilirubin (I-index) or of added intralipid (L-index) (SI units), according to manufacturer´s instructions. Alinity c utilizes the same wavelengths as Architect c8000/ c16000 (H-index: 500/524 nm; I-index: 572/604 nm and 628/660 nm; L-index: 524/804 nm) (15). Total bilirubin concentration and HIL indices were externally controlled once time per month using the Biochemistry Serum and Serum Indices Programmes, respectively, provided by The External Quality Control Programme from the Spanish Society of Laboratory Medicine (SEQC-ML).…”
Section: Methodsmentioning
confidence: 99%