2002
DOI: 10.1093/clinchem/48.3.566
|View full text |Cite
|
Sign up to set email alerts
|

Preanalytical Factors in the Measurement of Intact Parathyroid Hormone with the DPC IMMULITE Assay

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
12
0

Year Published

2003
2003
2024
2024

Publication Types

Select...
4
4

Relationship

0
8

Authors

Journals

citations
Cited by 24 publications
(12 citation statements)
references
References 4 publications
0
12
0
Order By: Relevance
“…Intact parathyroid hormone (PTH) was measured in EDTA plasma by immunochemiluminescence on the Immulite 2000 platform, ensuring complete sample filling of the collection vessel. 27,28 Interassay CV for intact PTH was 9% at 3AE2 pmol/l, 5AE6% at 30AE3 pmol/l and 5% at 104 pmol/l. 25OHD was measured by radioimmunoassay (RIA) using the DiaSorin RIA method with interassay CV of <10% at 14 and 33 nmol/l.…”
Section: Methodsmentioning
confidence: 93%
See 1 more Smart Citation
“…Intact parathyroid hormone (PTH) was measured in EDTA plasma by immunochemiluminescence on the Immulite 2000 platform, ensuring complete sample filling of the collection vessel. 27,28 Interassay CV for intact PTH was 9% at 3AE2 pmol/l, 5AE6% at 30AE3 pmol/l and 5% at 104 pmol/l. 25OHD was measured by radioimmunoassay (RIA) using the DiaSorin RIA method with interassay CV of <10% at 14 and 33 nmol/l.…”
Section: Methodsmentioning
confidence: 93%
“…Intact parathyroid hormone (PTH) was measured in EDTA plasma by immunochemiluminescence on the Immulite 2000 platform, ensuring complete sample filling of the collection vessel 27,28 . Interassay CV for intact PTH was 9% at 3·2 pmol/l, 5·6% at 30·3 pmol/l and 5% at 104 pmol/l.…”
Section: Methodsmentioning
confidence: 99%
“…This effect is due to the relatively high EDTA concentration within the sample mix, which chelates ions that are critical for the alkaline phosphatase indicator enzyme used in this assay. 19 These issues need to be appreciated by clinician users of this intact PTH assay, as well as by laboratories that offer this diagnostic test. Autoantibody and heterophile antibody interference is a recognized cause of diagnostic misclassification for all immunoassays, however this is thankfully rarely seen with intact PTH assays 20 (Table 3).…”
Section: Recommendations For Australian Laboratoriesmentioning
confidence: 99%
“…Insufficient sample volume (underfilling the vessel by ≥50%) can result in falsely low intact PTH values using the DPC Immulite assay. This effect is due to the relatively high EDTA concentration within the sample mix, which chelates ions that are critical for the alkaline phosphatase indicator enzyme used in this assay 19 . These issues need to be appreciated by clinician users of this intact PTH assay, as well as by laboratories that offer this diagnostic test.…”
Section: Diagnostic Criteria Of Phptmentioning
confidence: 99%
“…Each one used its routine laboratory kit that is used to measure intact PTH in dogs and cats (Laboratorio Echevarne, Calle Provenza, 312 Bajos, 08037 Barcelona, Spain) (Laboklin GmbH & Co. KG, Steubenstraße 4, 97688 Bad Kissingen, Germany). Blood on EDTA of two of the 8 guinea pigs randomly selected was sent at room temperature and analyzed in less than 24 h by one laboratory (Laboratorio Echevarne, Calle Provenza, 312 Bajos, 08037 Barcelona, Spain), because the PTH levels in these samples are proven to be higher than in frozen serum [ 7 , 15 , 16 ]. This analysis was performed to partially discard the influence of temperature and EDTA on the results obtained in frozen plasma.…”
Section: Methodsmentioning
confidence: 99%