1999
DOI: 10.1006/meth.1999.0812
|View full text |Cite
|
Sign up to set email alerts
|

Practical Confocal Microscopy and the Evaluation of System Performance

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

3
104
0
2

Year Published

2000
2000
2016
2016

Publication Types

Select...
7

Relationship

3
4

Authors

Journals

citations
Cited by 55 publications
(109 citation statements)
references
References 3 publications
3
104
0
2
Order By: Relevance
“…Regular system checks should be common practice and measure e.g. the point spread function (PSF), the laser power, the illumination homogeneity of the field of view, the detector sensitivity as signal‐to‐noise‐ratio (SNR) and the coefficient of variation (Cole et al, 2011, 2013; Gelman and Rietdorf, 2010; Hibbs, 2006; Stack et al, 2011; Theer et al, 2014; Zucker and Price, 2001, 1999). When performed regularly, standardized test series often help identify damaged and misaligned microscope components, even before they significantly impair the quality of user data.…”
Section: How To Design Set Up and Run Your Imaging Core Facilitymentioning
confidence: 99%
“…Regular system checks should be common practice and measure e.g. the point spread function (PSF), the laser power, the illumination homogeneity of the field of view, the detector sensitivity as signal‐to‐noise‐ratio (SNR) and the coefficient of variation (Cole et al, 2011, 2013; Gelman and Rietdorf, 2010; Hibbs, 2006; Stack et al, 2011; Theer et al, 2014; Zucker and Price, 2001, 1999). When performed regularly, standardized test series often help identify damaged and misaligned microscope components, even before they significantly impair the quality of user data.…”
Section: How To Design Set Up and Run Your Imaging Core Facilitymentioning
confidence: 99%
“…objective lens, averaging, pinhole size, bleaching, PMT voltage, laser excitation source, and spectral registration). In addition, the CLSM has to be checked and evaluated to ensure that it is delivering its proper performance (24)(25)(26)(27)(28)(29). Image quality is primarily dependent on the sample preparation.…”
Section: Discussionmentioning
confidence: 99%
“…The confocal microscope was tested for field illumination alignment, optical efficiency, colocalization, and axial resolution prior to embryo, fetal tissue, and insect observation (24)(25)(26). The lens was inspected and cleaned prior to use.…”
Section: Laser Scanning Confocal Microscopymentioning
confidence: 99%