Inoculation of an automated system for rapid identification (ID) and antimicrobial susceptibility testing (AST) directly from positive blood culture bottles will reduce the turnaround time of laboratory diagnosis of septicemic patients, which benefits clinical outcome and decreases patient costs. Direct test results, however, must always be confirmed by testing a pure overnight culture, which is the "gold standard." We studied the accuracy of direct testing versus repeat testing in order to investigate the possibility of refraining from repeat testing. We also assessed the clinical risk of reporting results based on direct testing only. We inoculated Vitek 2 (bioMérieux) directly from 410 positive BACTEC 9240 (BD) blood culture bottles containing gram-negative rods and studied the ID and AST results. In a comparison of direct inoculation with the standard method, a total of 344 isolates of Enterobacteriaceae and Pseudomonas aeruginosa were tested, and 93.0% were correctly identified. Of the 39 (10.2%) samples that contained bacilli not identifiable by Vitek 2, only 1 gave a conclusive, correct result. The overall MIC agreement among 312 isolates was 99.2%, with 0.8% very major and 0.02% major error rates. Of only three (polymicrobial) samples, the direct susceptibility pattern would be reported to the clinician as too sensitive. Vitek 2 results obtained from direct inoculation of blood culture bottles containing gram-negative bacilli are safe enough for immediate reporting, provided that ID and AST are consistent. Repeat testing is not necessary, unless Gram stain or overnight subculture results raise doubt about the purity of the culture.Shortening the turnaround time of microbiological analyses with an automated system for rapid identification and susceptibility testing of bacteria leads to a significant reduction of patient morbidity, mortality, and cost (1, 2). In particular, for patients with septicemia, rapid laboratory results are essential for appropriate treatment and improving clinical outcome (21). An automated blood culture system that monitors culture bottles for microbial growth minimizes the time necessary to detect positive blood cultures. Another way to save time might be to inoculate an automated system for rapid identification and susceptibility testing directly from positive blood culture bottles. The conventional way is to wait for the overnight subculture on agar to prepare a standard inoculum according to the manufacturer's guidelines.Of all bacteria involved in bloodstream infections, the fastgrowing Enterobacteriaceae will probably produce the quickest and best-correlating results for direct and standard inoculation. Several studies have compared direct and standard methods for different (combinations of) automated systems, but to our knowledge no study has yet done this for Vitek 2 (7,8,18,20,22).We routinely inoculate Vitek 2 directly from BACTEC 9240 blood culture bottles that are positive for gram-negative rods. To check the direct results, we repeat the tests the next day with suspe...