Poster — Wed Eve—02: Monte Carlo Dose Calculation of Critical Organs in MDR Cs‐137 Afterloading Intracavitary Brachytherapy

Abstract: The aim of this study is to perform Monte Carlo dose calculations on two critical organs, bladder and rectum, by considering the effects of applicator attenuation for three different applicator diameters of the Selectron/MDR intracavitary brachytherapy system. Three different pelvic phantoms, with a gynecological cylinder, were used to simulate the treatments. Dosimetric measurements were carried out in five different positions in the rectum and one position in the bladder using p‐type diode detectors. The abs… Show more