Possible Use of a Safety-Valve with a Foley Catheter During Catheterisation of Male Spinal Cord Injury Patients for Prevention of Urethral Trauma Caused by Inflation of the Catheter Balloon in the Urethra
Abstract:We used a safety-valve (Trans-Urethral Catheterisation Safety Valve, Class Medical, Limerick, Ireland) to prevent urethral trauma due to inflation of the anchoring balloon in the urethra during catheterisation of male spinal cord injury patients in a spinal unit. The safety-valve is attached to the balloon channel of a Foley catheter. If the balloon is inflated when it is in the urethra, the pressure valve is activated. Any fluid pushed into the balloon channel leaks out and balloon inflation stops, indicatin… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.