2000
DOI: 10.1302/0301-620x.82b2.0820185
|View full text |Cite
|
Sign up to set email alerts
|

Poor results of the Freeman uncemented metal-backed acetabular component

Abstract: The Freeman metal-backed acetabular component is specifically designed for use with a Freeman-neck-retaining stem.1 Among its several unique features are an arc of 140°; subtended in relation to the prosthetic femoral head and a titanium metal back with a superolateral flange through which a screw can be inserted to supplement uncemented press-fit fixation. Into this metal back is inserted a high-density polyethylene (HDP) liner which has a central peg and two superolateral lugs that lock into the metal back (… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

0
2
0

Year Published

2003
2003
2007
2007

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(2 citation statements)
references
References 7 publications
0
2
0
Order By: Relevance
“…A similar problem was noted by Harris 21 In primary replacement, a similar acetabular implant has given disappointing results with survival ranging from 69% at five years 23 to 83% at ten years. 24 The press-fit nature of the polyethylene insert into the metal backing allows some micromovement between the respective parts. This provides a source of wear particles which produce an inflammatory reaction causing loosening.…”
Section: Discussionmentioning
confidence: 99%
“…A similar problem was noted by Harris 21 In primary replacement, a similar acetabular implant has given disappointing results with survival ranging from 69% at five years 23 to 83% at ten years. 24 The press-fit nature of the polyethylene insert into the metal backing allows some micromovement between the respective parts. This provides a source of wear particles which produce an inflammatory reaction causing loosening.…”
Section: Discussionmentioning
confidence: 99%
“…[13][14][15][16] Other factors contributing to loosening of the cup include previous surgery, infection, impingement, and failure of the implant. [17][18][19][20] In addition to surgical technique, the physical properties of the material and its surface structure play an important role in loosening of the cup. Our study focused on the different incidences of the radiological signs of loosening between various cup systems over a long period of time.…”
Section: Discussionmentioning
confidence: 99%