2008
DOI: 10.1373/clinchem.2008.105478
|View full text |Cite
|
Sign up to set email alerts
|

Pitfalls and Prevention Strategies for Liquid Chromatography-Tandem Mass Spectrometry in the Selected Reaction– Monitoring Mode for Drug Analysis

Abstract: BACKGROUND:We observed cases of false-positive results with the use of liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS). Different LC-MS/MS techniques that use the selected reaction-monitoring mode, routinely employed for the analysis and quantification of drugs and toxic compounds in biological matrices, were involved in the false-positive and potentially false-positive results obtained. We sought to analyze the causes of and solutions to this problem.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

2
41
0
1

Year Published

2009
2009
2020
2020

Publication Types

Select...
5
5

Relationship

3
7

Authors

Journals

citations
Cited by 97 publications
(44 citation statements)
references
References 30 publications
2
41
0
1
Order By: Relevance
“…Noticeably, the EPI spectrum of quetiapine includes major fragments at m/z 279, 253, 247, and 221. These fragments followed the same fragmentation pathways as those we described for some phenothiazines in a recent paper [18]. Particularly, the loss of 32 amu between the fragments at m/z 279 and 247 corresponds to the loss of S from the cyclic structure.…”
Section: Case 1: Identification Of Trazodone Metabolites In a Forensisupporting
confidence: 64%
“…Noticeably, the EPI spectrum of quetiapine includes major fragments at m/z 279, 253, 247, and 221. These fragments followed the same fragmentation pathways as those we described for some phenothiazines in a recent paper [18]. Particularly, the loss of 32 amu between the fragments at m/z 279 and 247 corresponds to the loss of S from the cyclic structure.…”
Section: Case 1: Identification Of Trazodone Metabolites In a Forensisupporting
confidence: 64%
“…Our assay is based on the specific quantification of CN-dependent dephosphorylation of a synthetic phosphopeptide substrate by LC-MS/MS in MRM mode, a technique widely available in clinical laboratories (17 ). The use of a stable isotopelabeled form of RII peptide, as an internal standard, helps to reduce experimental variation during sample preparation and acquisition.…”
Section: Discussionmentioning
confidence: 99%
“…Ainsi, trois pics supplémentaires importants ont été repérés correspondant pour l'un à une perte d'un groupement méthyle et pour les deux autres à la perte d'un groupement méthyle, plus l'ajout d'un oxygène dans la formule de l'alimé-mazine [32]. L'enrichissement de base de données de référence, par l'ajout de paramètres caractérisant quelques métabolites principaux de xénobiotiques a plusieurs intérêts majeurs [33].…”
Section: Discussionunclassified