Pharyngeal and esophageal pressure measurements to evaluate respiratory mechanics in infants on high flow nasal cannula: A feasibility study

Abstract: High-flow humidified nasal cannula (HFNC) is often\ud used to relieve respiratory distress in children with acute\ud pulmonary disease, although its effects on respiratory mechanics\ud have not been objectively studied. The purpose of this study was\ud to test the feasibility of measuring pharyngeal (PP) and\ud esophageal (Pes) pressures of young children on HFNC oxygen\ud therapy through a specifically designed new monitoring,\ud acquisition, and elaboration system (MAES). Through MAES we\ud recorded and elab… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
3
0

Year Published

2016
2016
2024
2024

Publication Types

Select...
5
1

Relationship

2
4

Authors

Journals

citations
Cited by 6 publications
(5 citation statements)
references
References 14 publications
0
3
0
Order By: Relevance
“…Flow was detected using a pneumotachograph (4,500 series, not heated, flow range 0–35 LPM; Hans Rudolph, Shawnee, Kansas) combined with a differential pressure transducer (SensorTechnics 144LU01D-PCB, SensorTechnics, Inc., Mansfield, CA), inserted into the facemask port. Pes, as a surrogate for pleural pressure, was measured with a single 1.3 mm (approximately 4 Ch) diameter solid-state pressure transducer at the catheter tip (CTO-1, Gaeltec, Dunvegan, Scotland) ( 23 ). Correct catheter placement in the lower third of the esophagus was confirmed using the occlusion technique ( 24 ).…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…Flow was detected using a pneumotachograph (4,500 series, not heated, flow range 0–35 LPM; Hans Rudolph, Shawnee, Kansas) combined with a differential pressure transducer (SensorTechnics 144LU01D-PCB, SensorTechnics, Inc., Mansfield, CA), inserted into the facemask port. Pes, as a surrogate for pleural pressure, was measured with a single 1.3 mm (approximately 4 Ch) diameter solid-state pressure transducer at the catheter tip (CTO-1, Gaeltec, Dunvegan, Scotland) ( 23 ). Correct catheter placement in the lower third of the esophagus was confirmed using the occlusion technique ( 24 ).…”
Section: Methodsmentioning
confidence: 99%
“…Flow and Pes signals were sampled simultaneously using a portable oscilloscope (PicoScope 4,824, Pico Technology, UK), and acquired for off-line analysis ( 25 ). Flow and Pes tracings were combined to obtain data for respiratory mechanics (dynamic compliance, Cdyn, and inspiratory and expiratory airway resistances, Rinsp, Rexp, respectively) and respiratory effort (esophageal pressure-time product per minute, PTPes/min) indices using specific routines (MATLAB® software), as previously reported ( 23 , 25 ).…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…For this purpose, based on our previous experience [11], we specifically designed a new programmable data monitoring, acquisition, and elaboration system (Pro_HFNC) and tested its effectiveness on a series of young infants hospitalized in the Pediatric Intensive Care Unit (PICU) at Sapienza University of Rome (Italy) with mild to moderate acute respiratory failure treated with HFNC.…”
Section: Introductionmentioning
confidence: 99%
“…Giving HFNC to COVID-19 patients starts in the flow range of 30-50 LPM with an oxygen concentration of 92% [15]. The use of HFNC should be strictly monitored by nurses because when the patient is in the same condition or worsens the nurse is required to increase the flow rate given to the patient, this monitoring is carried out every hour so that nurses can know the results of giving HFNC and if the patient's condition remains then the flow is given to HFNC will be upgraded [16] [17], Therefore, during the administration of oxygen therapy using HFNC it must be monitored carefully, to monitor the dose of oxygen administration to the patient does not decrease or does not stop. [18].…”
Section: Introductionmentioning
confidence: 99%