PET Demonstrator for a Human Brain Scanner Based on Monolithic Detector Blocks

Abstract: We have implemented and evaluated a positron emission tomography (PET) demonstrator using two monolithic detector blocks operating in coincidence with dedicated applicationspecific integrated circuit (ASIC) readout. Each detector is composed of a monolithic lutetium yttrium orthosilicate (LYSO) scintillator coupled to a pair of Hamamatsu S8550-02 APD arrays. The front-end electronics of this demonstrator is based on the VATA240 ASIC readout, which sums the charge provided by each row and column of the APD arra… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
18
0
2

Year Published

2011
2011
2022
2022

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 15 publications
(21 citation statements)
references
References 19 publications
1
18
0
2
Order By: Relevance
“…For PET applications, multiple tests using APD detectors were performed [38,39,40,41,42,43], showing increased acceptance and support for APD-based scanners over PMTs moving forward. Although the gain of APDs was not as good as in PMTs, they offered more benefits such as reduced cost, smaller size, and insensitivity to magnetic fields.…”
Section: Avalanche Photodiodes (Apds)mentioning
confidence: 99%
See 1 more Smart Citation
“…For PET applications, multiple tests using APD detectors were performed [38,39,40,41,42,43], showing increased acceptance and support for APD-based scanners over PMTs moving forward. Although the gain of APDs was not as good as in PMTs, they offered more benefits such as reduced cost, smaller size, and insensitivity to magnetic fields.…”
Section: Avalanche Photodiodes (Apds)mentioning
confidence: 99%
“…Another such organ-specific project was a brain PET [43] insert for existing MRI equipment. Compared to dual-modality PET/CT systems, PET/MRI systems can reduce the risk of ionizing radiation, provide better soft tissue contrast of grey and white matter, and correct for attenuation and scattering of PET images [43]. The building block of this scanner consisted of large, dual layer monolithic LYSO:Ce scintillators of trapezoidal shape coupled to a pair of Hamamatsu S8550-02 APD arrays.…”
Section: Avalanche Photodiodes (Apds)mentioning
confidence: 99%
“…We have developed a simple PET demonstrator for evaluating the performance of the FlexToT ASIC in the readout of PET detector blocks, similar to one previously used by our group at CIEMAT [9]. The demonstrator consists of two detector boxes aligned face to face, each containing a set of FlexToT front-end readout boards with scintillators mounted on the SiPM matrices, at a face-to-face nominal distance of 200 mm.…”
Section: E Pet Demonstrator Setupmentioning
confidence: 99%
“…On the other hand, insufficient sensitivity results in a low signal‐to‐noise ratio (SNR), which restricts the capability of performing high‐resolution and dynamic imaging 32 . Although several dedicated brain PET scanners have been developed for research 33–37 as well as proof‐of‐principle simulations to assess their capabilities, 38–40 these systems still suffer from trade‐offs among cost, sensitivity, spatial resolution, and timing resolution.…”
Section: Introductionmentioning
confidence: 99%