Performance of a novel pressure release valve for cuff pressure control in pediatric tracheal tubes

Abstract: These preliminary results show that the CP pop-off valve limits the effect of rapid, potentially dangerous manual cuff inflation maneuvers and reliably prevents CP exceeding the predetermined level of 20 cmH(2)O when exposed to nitrous oxide.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
18
0
1

Year Published

2005
2005
2021
2021

Publication Types

Select...
8
1

Relationship

1
8

Authors

Journals

citations
Cited by 23 publications
(19 citation statements)
references
References 29 publications
(25 reference statements)
0
18
0
1
Order By: Relevance
“…The use of paediatric high volume-low pressure cuffs does not prevent high pressures being generated. Cuff pressure monitoring or cuff pressure limitation using a pressure release valve or cuff pressure regulator is therefore important [13]. Manual pilot balloon palpation has been shown to be unreliable to estimate cuff pressure [14].…”
Section: Discussionmentioning
confidence: 99%
“…The use of paediatric high volume-low pressure cuffs does not prevent high pressures being generated. Cuff pressure monitoring or cuff pressure limitation using a pressure release valve or cuff pressure regulator is therefore important [13]. Manual pilot balloon palpation has been shown to be unreliable to estimate cuff pressure [14].…”
Section: Discussionmentioning
confidence: 99%
“…Since cuff palpation has been shown to be an unreliable method of monitoring cuff pressure, cuff inflation should be undertaken using cuff pressure monitoring or by using an automatic cuff pressure release device [7,20,[22][23][24]. Increasing the volume of the cuff pilot balloon is an additional measure to increase compliance and the residual inflation volume.…”
Section: Discussionmentioning
confidence: 99%
“…[33,48,[55][56][57]. The Microcuff PET is provided with a pediatric cuff manometer and a cuff pressure release valve which limits the cuff pressure at about 20 cm H 2 O (FIGURE 4) [58].…”
Section: Package With Tube and Patient-related Informationmentioning