2018
DOI: 10.1186/s40658-018-0229-4
|View full text |Cite
|
Sign up to set email alerts
|

Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system

Abstract: BackgroundPoleStar m660 is a newly developed clinical PET/CT system with time-of-flight (TOF) capability. The aim of this study is to characterize the performance of the new system. Spatial resolution, sensitivity, scatter fraction, and noise equivalent count rate (NECR) were measured on the scanner according to the NEMA NU 2-2012 protocol. The timing resolution was measured using a rotating line source that orbited around the center of field of view (CFOV) at a radius of 20 cm. The image quality phantom was a… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
16
0

Year Published

2018
2018
2024
2024

Publication Types

Select...
8

Relationship

2
6

Authors

Journals

citations
Cited by 15 publications
(16 citation statements)
references
References 17 publications
(17 reference statements)
0
16
0
Order By: Relevance
“…Currently, a clinical evaluation of the whole-body PoleStar m660 TOF-PET/CT scanner is carried out [ 51 ]. The system is based on 3.63×3.63×20 mm 3 LYSO scintillators arranged into detector modules (14×14 crystals).…”
Section: Reviewmentioning
confidence: 99%
“…Currently, a clinical evaluation of the whole-body PoleStar m660 TOF-PET/CT scanner is carried out [ 51 ]. The system is based on 3.63×3.63×20 mm 3 LYSO scintillators arranged into detector modules (14×14 crystals).…”
Section: Reviewmentioning
confidence: 99%
“…All imaging was performed on a clinical PET-CT system [ 18 ] (PoleStar m660, Sinounion Healthcare Inc., Beijing, China) at PUMCH. 11 C-acetate was synthesized as described in literature [ 19 , 20 ].…”
Section: Methodsmentioning
confidence: 99%
“…According to the Figure 9: Setup for the detection efficiency measurement. From left to right : the PbF 2 detection module (black squared block) with amplifiers PCB (green plate); the 22 Na radioactive source (transparent plastic disk); YAP spectrometer (black cylinder with the metallic ring).…”
Section: Efficiency Measurementmentioning
confidence: 99%
“…Measured and simulated spectra, produced by a 22 Na radioactive source in the YAP spectrometer. The pink histogram is the measured spectrum.…”
Section: Figure 10mentioning
confidence: 99%