Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT

Abstract: Background This paper describes the National Electrical Manufacturers Association (NEMA) system performance of the Discovery MI 3-ring PET/CT (GE Healthcare) installed in Bruges, Belgium. This time-of-flight (TOF) PET camera is based on silicon photomultipliers instead of photomultiplier tubes. Methods The NEMA NU2-2012 standard was used to evaluate spatial resolution, sensitivity, image quality (IQ) and count rate curves of the system. Timing and energy resolution were… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

2
32
0

Year Published

2019
2019
2022
2022

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 30 publications
(34 citation statements)
references
References 27 publications
2
32
0
Order By: Relevance
“…It is noteworthy that the results for the two systems are similar with, for instance, a spatial resolution for DMI and DMI-DR of 4.62 vs 4.58 for FBP, 3.85 vs 4.13 for VPHD (3D OSEM) and 2.70 vs 2.84 for VPHD-S (3D OSEM + PSF) at 1-cm radial offset. For the DMI evaluation, the results obtained at our institution are compared with previous results from Stanford and Uppsala Universities [2], Tokyo [3] and Brugge [4] in Fig. 1a-c.…”
Section: Spatial Resolutionmentioning
confidence: 99%
See 3 more Smart Citations
“…It is noteworthy that the results for the two systems are similar with, for instance, a spatial resolution for DMI and DMI-DR of 4.62 vs 4.58 for FBP, 3.85 vs 4.13 for VPHD (3D OSEM) and 2.70 vs 2.84 for VPHD-S (3D OSEM + PSF) at 1-cm radial offset. For the DMI evaluation, the results obtained at our institution are compared with previous results from Stanford and Uppsala Universities [2], Tokyo [3] and Brugge [4] in Fig. 1a-c.…”
Section: Spatial Resolutionmentioning
confidence: 99%
“…Figure 2c, d presents the results obtained for the DMI system at Stanford and Uppsala Universities [2], Tokyo [3] and Brugge [4] in comparison with the results obtained at our institution using the VPFX algorithm. It is noteworthy that image quality results are very similar across centers, apart from the BV obtained at Brugge for the [2], Tokyo [3] and Brugge [4] measurements for DMI (c-d) and to Southampton-Poole-Plymouth [5] for DMI-DR (e-f) 3-ring system, with a minimum/maximum relative deviation of 74%/50% from our results is obtained. Figure 2e, f shows similar image quality results obtained at Southampton-Poole-Plymouth [5] and our center for the DMI-DR system.…”
Section: Image Qualitymentioning
confidence: 99%
See 2 more Smart Citations
“…Scans were performed on a TOF 3-ring Discovery 710 PET/CT scanner with lutetium yttrium orthosilicate (LYSO) crystals (GE Healthcare), see [21] for an overview of the performance of the system. Images were reconstructed using 3 different algorithms, each using low dose CT information for attenuation correction and the same normalization correction factors with scatter and randoms corrected.…”
Section: Pet/ct Scanning and Reconstructionmentioning
confidence: 99%