2014
DOI: 10.4137/cmc.s15229
|View full text |Cite
|
Sign up to set email alerts
|

Percutaneous Vascular Closure System Based on an Extravascular, Bioabsorbable Polyglycolic Plug (ExoSeal): Results from 1000 Patients

Abstract: PURPOSEThe objective of this prospective study was to evaluate the effectiveness and safety of a percutaneous closure system based on a polyglycolic acid plug for achieving hemostasis.MATERIAL AND METHODSIn this study from 2011 to 2014, a percutaneous vascular closure system (ExoSeal) was used in 1000 patients (mean age 70.6 ± 10.2 years), using antegrade and retrograde techniques within the context of an angiographic intervention. The system was used in conjunction with transfemoral approaches with a sheath s… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

0
2
0

Year Published

2018
2018
2021
2021

Publication Types

Select...
3
2

Relationship

0
5

Authors

Journals

citations
Cited by 6 publications
(2 citation statements)
references
References 20 publications
0
2
0
Order By: Relevance
“…Previous reports in ExoSeal series with 6 Fr access have found minor complication rates of 1-1.8% for pseudoaneurysm and 3.4-6.3% for inguinal hematoma. [11][12][13] Thus, sheath size may influence the access site complication rate. In addition, Ramirez et al 14 found that closure device use with CFA antegrade access had a higher risk than retrograde access (any access site hematoma, odds ratio = 1.46).…”
Section: Discussionmentioning
confidence: 99%
“…Previous reports in ExoSeal series with 6 Fr access have found minor complication rates of 1-1.8% for pseudoaneurysm and 3.4-6.3% for inguinal hematoma. [11][12][13] Thus, sheath size may influence the access site complication rate. In addition, Ramirez et al 14 found that closure device use with CFA antegrade access had a higher risk than retrograde access (any access site hematoma, odds ratio = 1.46).…”
Section: Discussionmentioning
confidence: 99%
“…Kamusella et al reported a case of device related stenosis (0.1%) among 1000 patients in whom the EXOSEAL VCD was used. The stenosis was detected by the partially intravascular position of the plug, but hemodynamically significant stenosis was not evident [ 4 ]. As described above, complications of arterial occlusion induced by EXOSEAL VCD are thought to be very rare.…”
Section: Discussionmentioning
confidence: 99%