Abstract:Urological Survey review was performed to retrieve data on safety and efficacy, complications and outcomes using this product. Results: Six patients (17%) to date have presented with defective vaginal healing manifested by extrusion of the sling material. Mean time to presenting symptoms was 9 months (range 2 to 15). All patients required surgical removal of the sling material. No urethral erosions were noted. Conclusions: Our experience suggests that the IVS sling system, which uses a multi-filament polypropy… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.