Abstract:Results Average MAFA-Ratio values calculated from pre-and post-stent placement were significantly lower after deployment of the 64-wire device (mean= 0.62±0.09) compared to the 48-wire device (0.71±0.06); p=<0.05. Conclusions Our in-vitro results show that the 64 wire FDS (Evolve) had superior flow diversion effect compared to the 48 wire FDS (Pipeline), suggesting that 64-wires are superior to 48-wire designs for flow diversion efficacy. Disclosures N.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.