Overall performance of Bovine Corneal Opacity and Permeability (BCOP) Laser Light-Based Opacitometer (LLBO) test method with regard to solid and liquid chemicals testing

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
17
0
1

Year Published

2021
2021
2024
2024

Publication Types

Select...
4
2

Relationship

0
6

Authors

Journals

citations
Cited by 7 publications
(21 citation statements)
references
References 14 publications
0
17
0
1
Order By: Relevance
“…The methods used in DAL-1 and DAL-2 encompass the following validated test methods: the Bovine Corneal Opacity and Permeability (BCOP) using the laser light-based opacitometer (LLBO) 1 according to the OECD TG 437, the Reconstructed human Cornea-like Epithelium (RhCE) according to the OECD TG 492 2 and the Short Time Exposure in vitro (STE) according to the OECD TG 491. Transferability, within-and between-laboratory reproducibility of these individual test methods have been assessed during their respective validation studies (16,17,18,19,20).…”
Section: Das and Use Scenarios Included In The Guidelinementioning
confidence: 99%
See 1 more Smart Citation
“…The methods used in DAL-1 and DAL-2 encompass the following validated test methods: the Bovine Corneal Opacity and Permeability (BCOP) using the laser light-based opacitometer (LLBO) 1 according to the OECD TG 437, the Reconstructed human Cornea-like Epithelium (RhCE) according to the OECD TG 492 2 and the Short Time Exposure in vitro (STE) according to the OECD TG 491. Transferability, within-and between-laboratory reproducibility of these individual test methods have been assessed during their respective validation studies (16,17,18,19,20).…”
Section: Das and Use Scenarios Included In The Guidelinementioning
confidence: 99%
“…2. However, previous data(20) showed that out of 22 in vivo No Cat. liquids, no liquid was mispredicted as Cat.…”
mentioning
confidence: 92%
“…1 by the BCOP LLBO. This assumption is based on BCOP LLBO data that were published by Adriaens et al 14 In total, BCOP LLBO data were available for 46 in vivo No Cat. chemicals (22 liquids and 24 solids).…”
Section: Performance Of Dal-1 and Dal-2 By Un Ghsmentioning
confidence: 99%
“…chemicals (22 liquids and 24 solids). 14 Only 1/46 (2.2%) chemicals, a solid, was predicted Cat. 1 based on the BCOP LLBO.…”
Section: Performance Of Dal-1 and Dal-2 By Un Ghsmentioning
confidence: 99%
“…When test chemicals within certain chemical (i.e., alcohols, ketones) or physical (i.e., solids) classes are excluded from the database, the BCOP OP-KIT test method has an overall accuracy of 85% (n=131), a false positive rate of 20% (n=81), and a false negative rate of 8% (n=50) for the UN GHS classification system (3). The BCOP LLBO test method has an overall accuracy of 78% (n=145), a false positive rate of 21% (n=89), and a false negative rate of 24% (n=56), when compared to in vivo rabbit eye test method data classified according to the UN GHS classification system (12) (see Annex 2, Table 3). When test chemicals within certain chemical (i.e., alcohols, ketones) or physical (i.e., solids) classes are excluded from the database, the BCOP LLBO test method has an overall accuracy of 85% (n=55), a false positive rate of 17% (n=32), and a false negative rate of 13% (n=23) for the UN GHS classification system (12).…”
mentioning
confidence: 99%