2019
DOI: 10.1002/mp.13894
|View full text |Cite
|
Sign up to set email alerts
|

Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations — perpendicular and parallel

Abstract: The goal of the present work was to provide a large set of detector-specific output correction factors for seven small volume ionization chambers on two linear accelerators in four megavoltage photon beams utilizing perpendicular and parallel orientation of ionization chambers in the beam for nominal field sizes ranging from 0.5 cm 2 9 0.5 cm 2 to 10 cm 2 9 10 cm 2. The present study is the second part of an extensive research conducted by our group. Methods: Output correction factors k f clin ;f ref Q clin ;Q… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
21
1

Year Published

2019
2019
2023
2023

Publication Types

Select...
6
1

Relationship

1
6

Authors

Journals

citations
Cited by 22 publications
(22 citation statements)
references
References 38 publications
0
21
1
Order By: Relevance
“…42 determined CFs exhibit a difference of up to 4% for 0.5 × 0.5 cm 2 field, although for 10 MV beam this difference decreases to approximately 2.5 %. For PTW31023, the difference in CF measured in this study and that of Casar et al 24 is <1.5% for fields smaller than 1 × 1 cm 2 for both energies. Whereas a substantial deviation has been observed for smaller fields.…”
Section: Discussioncontrasting
confidence: 69%
See 2 more Smart Citations
“…42 determined CFs exhibit a difference of up to 4% for 0.5 × 0.5 cm 2 field, although for 10 MV beam this difference decreases to approximately 2.5 %. For PTW31023, the difference in CF measured in this study and that of Casar et al 24 is <1.5% for fields smaller than 1 × 1 cm 2 for both energies. Whereas a substantial deviation has been observed for smaller fields.…”
Section: Discussioncontrasting
confidence: 69%
“…However, It has been characterized for reference and relative dosimetry in the conventional fields by Bushing et al and Bruggmoser et al 22,23 Recently, it has also been studied by Casar et al for determination of CFs in radial and axial orientations. 24 However, the response has not been corrected for polarity and recombination effects as recommended by TRS483. Similarly, RNC, the smallest commercially available IC, has been characterized for different dosimetric parameters specially the polarity and stem effects.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…It requires determination of detector specific correction factors, which have to be utilized individually for the selected detector and are associated with additional uncertainties. 24 , 25 , 26 The latter can pose a problem to conduct such a study with sufficient reliability and robustness.…”
Section: Discussionmentioning
confidence: 99%
“…In 2019, Francescon et al [121] investigated the sensitivity of dosimetric correction factors to interunit variation and reported variation up to 9% between the measured data corrected using the recommendations of TRS-483 and Monte Carlo results. Casar et al [122] provided detector-specific output correction factors for small-field sizes using the recommendations of TRS-483.…”
Section: Monte Carlo Studies For Small-field Dosimetrymentioning
confidence: 99%