Abstract:Objectives
This study aimed to evaluate the influence of different CBCT acquisition protocols on reducing effective radiation dose while maintaining image quality.
Materials and Methods
The effective dose emitted by a CBCT Picasso Trio® device was calculated using thermoluminescent dosimeters (TLDs) placed in a Rando Alderson phantom, following the ICRP 103 (2007) guidelines. Image quality was assessed by three experienced evaluators, who examined CBCT images for anatomical structure identification and image… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.