Optimization of Graphite Furnace Atomic Absorption Spectrophotometry for Determination of Trace Cadmium, Lead and Nickel in Urine

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
3
2

Year Published

1994
1994
2011
2011

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 12 publications
(5 citation statements)
references
References 20 publications
0
3
2
Order By: Relevance
“…The values of Pb in all three biological samples of steel mill paralysis workers are higher compared to the results obtained by other workers in the same field [65]. The exposed mean value of lead in urine samples was 58.8±21.3 μg/L as reported by Lin [66] and 71.7±26.3 μg/L as reported by Burguera et al [67]. These values are lower than our results (p<0.01).…”
Section: Discussioncontrasting
confidence: 87%
“…The values of Pb in all three biological samples of steel mill paralysis workers are higher compared to the results obtained by other workers in the same field [65]. The exposed mean value of lead in urine samples was 58.8±21.3 μg/L as reported by Lin [66] and 71.7±26.3 μg/L as reported by Burguera et al [67]. These values are lower than our results (p<0.01).…”
Section: Discussioncontrasting
confidence: 87%
“…The values of Pb in all three biological samples of steel mill workers are higher compared to the results obtained by other workers in the same field (Pan et al, 1993). The exposed mean value of lead in urine samples were 58.89/21.3 mg/L as reported by Lin (1991) and 71.79/26.3 mg/L, as reported by Burguera et al, (1986). These values are lower than our results (PB/0.01).…”
Section: Resultscontrasting
confidence: 54%
“…The well known and widely used procedures for Ni determination in serum need prior protein precipitation, using nitric acid [15,16,17,18,19], Triton X-100 [17,18] or trichloracetic acid [1] and ETAAS measurement of Ni in protein-free supernatant. Experiments performed showed that direct ETAAS measurement of Ni in serum is almost impossible, due to very high values of background absorption.…”
Section: Etaas Measurement Of Ni In Serummentioning
confidence: 99%
“…Although direct spectrometric determination is therefore preferable, matrix interferences were seen as limiting factors in this case. In many papers some of these problems are overcome by using prior sample dilution with water [12], Triton X-100 [13,14], nitric acid [15,16] or by a mixture of nitric acid and Triton X-100 [10,17,18]. Chemical modifiers (NH 4 H 2 PO 4 [7,10] or Pd [2,5,6]) have also been recommended for Ni ETAAS determination in blood serum and urine samples.…”
Section: Introductionmentioning
confidence: 99%