In this work, we have devised a biosensor utilizing an aptamer functionalized screen-printed carbon electrode (SPCE), for the quantitative estimation of acephate, using cyclic voltammetry (CV). The amino-modified aptamer, specific for acephate, was covalently attached to the surface of SPCE, using the electrochemical diazonium coupling method, followed by end-group activation. The surface structure and electrochemical characteristics of the aptasensor were specified by the application of SEM, FTIR, and CV. Response of the devised aptasensor was linear with acephate concentration in the range of 10 x 10− 9 g mL− 1 to 160 x 10− 9 g mL− 1 (R² = 0.9332, n = 3), with a limit of detection (LOD) 4.34 x 10− 9 g mL− 1. The efficiency of the aptasensor was validated in spiked water samples, consequently, the recovery percentage, of 98.7–104.66% and RSD of 4.59–5.27% (n = 3), was achieved.