2024
DOI: 10.3390/jcm13030905
|View full text |Cite
|
Sign up to set email alerts
|

Optimal Plate Choice for High-Neck Mandibular Condyle Fracture: A Mechanistic Analysis of 16 Options

Jakub Okulski,
Marcin Kozakiewicz,
Michał Krasowski
et al.

Abstract: (1) Background: Mandibular fractures are common, with the condylar process being a frequent site of injury, accounting for 25–45% of cases. This research aims to assess the mechanical suitability of various plates for high-neck condyle fractures. (2) Methods: Polyurethane models mimicking high-neck condyle fractures were utilized in this study. Sixteen distinct plate designs, constructed from titanium sheets, were tested. The figures underwent force assessments on a durability testing apparatus, and the relati… Show more