Abstract:This study compares a standard monofocal intraocular lens (IOL) with two rotationally asymmetric refractive multifocal IOLs, analyzing power profiles and wavefront data across three nominal powers. Tested IOLs included monofocal Acunex AN6 and multifocal AN6V (addition +1.50 D) and AN6VM (addition +3.00 D) by Teleon Surgical, at powers +10.00 D, +20.00 D, and +30.00 D. Using NIMO TR1504, power profiles, spherical aberration, and coma were analyzed. Multifocal lenses displayed an asymmetrical design, with zones… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.