Abdominal entry is the critical initial step of laparoscopic surgery as it is associated with unexpected complications. Acceptable working space is a prerequisite for safe minimal access surgery. Direct trocar insertion (DTI) technique has been often reported to be time-consuming and capable of causing injuries. We aim to facilitate conventional DTI employing an easy-to-use self-designed abdominal wall-lifting device, OneShot-M. We tested its safety in a porcine model. 18 pigs were randomized into three groups (n=6): open laparoscopy (Group A), DTI (Group B), DTI using OneShot-M (Group C), suitable for a 5-mm trocar. Outcomes were access time, entry-related minor and major complications. Median access time was shorter in C (43 seconds; range: 35-49 seconds), followed by B (82.5; 72-90) and A (190.5; 182-240). Statistically significant differences in procedural time between techniques were detected (p-value = 0.0005). Pairwise comparisons exhibited a statistically significant difference between A and C in terms of access time. Only one difficulty in entry was seen in C; pigs experienced several occurrences in A and B. This pilot study suggests that OneShot-M appears to be a technically feasible, fast and safe method for pneumoperitoneum induction in a porcine model, as it provides sufficient exposure of the operative field without access-related complications. Further clinical studies are needed.