2018
DOI: 10.1016/j.ejmp.2018.06.101
|View full text |Cite
|
Sign up to set email alerts
|

[OA029] Optimization of phase space files from clinical linear accelerators

Abstract: surement period. However, with an external monitor chamber, the output variability was reduced to less than 0.05% (1 sd). For each beam quality, the relative standard deviation for the measured k Q factors for the seven ionization chambers was about 0.07%. Conclusions. The study demonstrates the feasibility of highly reproducible MV calibrations at the level of a secondary standards laboratory, and the closeness of agreement (sd < 0.1%) among the seven ionization chambers supports that generic k Q-factors may … Show more