Novel Size-Variable Dedicated Rodent Oxygenator for ECLS Animal Models: Introduction of the “Ratox” Oxygenator and Preliminary In-Vitro Results

Abstract: The survival rate of extracorporeal life support (ECLS) remains overall at 60 %. Research and development is slow, partly due to the lack of sophisticated experimental models. This publication introduces a dedicated rodent oxygenator (“RatOx”) and shows preliminary in vitro classification tests. The RatOx has an adaptable fiber module size for various rodent models. Gas transfer performance over the fiber module for different blood flows and fiber module sizes were tested according to DIN EN ISO 7199.At the ma… Show more