Laser Beam Radiometry 1988
DOI: 10.1117/12.944232
|View full text |Cite
|
Sign up to set email alerts
|

Nbs Laser Power And Energy Measurements

Abstract: The National Bureau of Standards (NBS) maintains a set of electrically calibrated calorimeters designed and built specifically for laser energy measurements.These calorimeters are used as national reference standards for the calibration of optical power and energy meters.NBS offers laser measurement services based on the standard calorimeters to the public at a variety of laser wavelengths and power ranges.The uncertainties associated with these measurements have recently been re-evaluated.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
5
0

Year Published

1994
1994
2008
2008

Publication Types

Select...
4
3

Relationship

0
7

Authors

Journals

citations
Cited by 12 publications
(6 citation statements)
references
References 1 publication
0
5
0
Order By: Relevance
“…The uncertainty associated with the absorbtivity of the standard calorimeters used in this service is discussed elsewhere by T. Scott. [15] 6.1.2 Electronics (Type B)…”
Section: Measurement Standard Uncertaintiesmentioning
confidence: 99%
See 1 more Smart Citation
“…The uncertainty associated with the absorbtivity of the standard calorimeters used in this service is discussed elsewhere by T. Scott. [15] 6.1.2 Electronics (Type B)…”
Section: Measurement Standard Uncertaintiesmentioning
confidence: 99%
“…The uncertainty for the inequivalence of the standard calorimeters used in this service is covered by T. Scott. [15] 6.1.5 Electrical Calibration (Type A)…”
Section: Iheater Leads (Type B)mentioning
confidence: 99%
“…The absolute calibration [ 7 ] of the test meter at the desired laser wavelength and power level P c [used in Eq. (3) ] can be done prior to or after the nonlinearity measurement, since they are two independent measurements.…”
Section: Measurement Procedures and Analysismentioning
confidence: 99%
“…Since the K-series instruments measure total energy, average laser power is found by dividing the total energy by the exposure time interval. This calorimeter has been in constant use as a primary standard over the past twenty-five years and has shown little or no change in response characteristics during this time [2]. The laboratory standard used in these comparison measurements, PM150-50C, is a commercial, water-cooled, thermopile-based probe with excellent repeatability and spatial uniformity.…”
Section: Nist Measurement Systemmentioning