Multicentre performance evaluation of the E170 Module for MODULAR ANALYTICS

Abstract: The E170 module was evaluated at 13 sites in an international multicentre study. The objective of the study was to assess the analytical performance of 49 analytes, and to collect feedback on the system's reliability and practicability. The typical, within-run coefficients of variation (CVs) for most of the quantitative assays ranged between 1 and 2% while a range of 2-4% was achieved with the infectious disease methods. Total precision CVs were found to be within the manufacturer's expected performance ranges… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
13
0

Year Published

2005
2005
2018
2018

Publication Types

Select...
7
1

Relationship

1
7

Authors

Journals

citations
Cited by 16 publications
(13 citation statements)
references
References 3 publications
0
13
0
Order By: Relevance
“…The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are removed by a washing step, and application of voltage to the electrode induces chemiluminescent emission, which is measured by a photomultiplier 15. According to the competitive assay design, an inverse proportional calibration curve is obtained.…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are removed by a washing step, and application of voltage to the electrode induces chemiluminescent emission, which is measured by a photomultiplier 15. According to the competitive assay design, an inverse proportional calibration curve is obtained.…”
Section: Methodsmentioning
confidence: 99%
“…Pretreated samples are combined with biotinylated analyte-specific antibodies and labeled analyte derivatives and recovered with streptavidin microparticles 15. Detection is by electrochemiluminescence (see Materials and Methods).…”
Section: Methodsmentioning
confidence: 99%
“…Nevertheless, this was unlikely to have influenced results because the two methods yield comparable results. 26 A third limitation was the use of plasma cortisol selectivity ratios as the reference standard. All AVS selectivity indices used in both research and clinical practice are arbitrary because they have not been formally linked to outcome data in an evidence-based manner.…”
Section: Hypertensionmentioning
confidence: 99%
“…Undiluted 100-l serum samples were eluted at 0.25 ml͞min with 10 mM Tris-buffered 140 mM NaCl containing 2 mM CaCl 2 , pH 8.0, monitored at 280 nm, and 250-l fractions were collected for CRP and cholesterol assay. Serum concentrations of estradiol and testosterone were measured by electrochemiluminescence immunoassay (Roche Diagnostics) (35) in the bleedout samples from the 56-week cohort.…”
mentioning
confidence: 99%