Abstract:Quality assurance Accuracy End-to-end test and the dose distribution from the radiochromic films. The global gamma indices of the measurement doses and the treatment planning system calculation doses were compared. Results: The large majority of the RT centers (6/8) fulfilled the dosimetric goals, with the measured and calculated doses at the specification points agreeing within ±3% for PTV and ±5% for OARS. At 2 centers, TPS underestimated the lung doses by about 6% and spinal cord doses by 8%. The mean perce… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.