2019
DOI: 10.1016/j.rpor.2018.11.005
|View full text |Cite
|
Sign up to set email alerts
|

Multi-institutional evaluation using the end-to-end test for implementation of dynamic techniques of radiation therapy in Thailand

Abstract: Quality assurance Accuracy End-to-end test and the dose distribution from the radiochromic films. The global gamma indices of the measurement doses and the treatment planning system calculation doses were compared. Results: The large majority of the RT centers (6/8) fulfilled the dosimetric goals, with the measured and calculated doses at the specification points agreeing within ±3% for PTV and ±5% for OARS. At 2 centers, TPS underestimated the lung doses by about 6% and spinal cord doses by 8%. The mean perce… Show more