2020
DOI: 10.1002/mp.14153
|View full text |Cite
|
Sign up to set email alerts
|

Multi‐institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance

Abstract: The SRS MapCHECK â , a recently developed patient-specific quality assurance (PSQA) tool for end-to-end testing of stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT), was evaluated in a multi-institution study and compared with radiochromic film. Methods: The SRS MapCHECK was used to collect data on 84 SBRT or SRS PSQA plans/fields at nine institutions on treatment delivery devices (TDD) manufactured by Varian and Elekta. PSQA plans from five different treatment planning software (T… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

9
78
0

Year Published

2020
2020
2022
2022

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 48 publications
(92 citation statements)
references
References 99 publications
9
78
0
Order By: Relevance
“…Many of these systems are unable to deliver true composite treatment plans so a failed QA measurement will do little to guide the practitioners on how to proceed and may potentially lead to looser QA passing tolerance levels. 14,15,17…”
Section: C Electronic Phantomsmentioning
confidence: 99%
See 3 more Smart Citations
“…Many of these systems are unable to deliver true composite treatment plans so a failed QA measurement will do little to guide the practitioners on how to proceed and may potentially lead to looser QA passing tolerance levels. 14,15,17…”
Section: C Electronic Phantomsmentioning
confidence: 99%
“…There are a number of electronic phantom devices that have been developed over the years and have had much success in the patient-specific verification of IMRT, VMAT 19 and more recently SRS. 15,17,18,20 These devices utilize either diodes or ion chambers to perform measurements and give immediate results. The immediacy of the results makes them an attractive option for routine quality assurance measurements.…”
Section: C Electronic Phantomsmentioning
confidence: 99%
See 2 more Smart Citations
“…The array has a measurable area of 77 mm × 77 mm. 12 The SRS MapCHECK can be inserted into the Sun Nuclear StereoPHAN which holds the SRS MapCHECK over the superior end of the couch. The StereoPHAN is a cylindrical PMMA phantom which is shaped in order to mimic the shape of a head.…”
Section: Introductionmentioning
confidence: 99%