Abstract: Background Occlusions of intravenous (IV) tubing can prevent vital and time-critical medication or solutions from being delivered into the bloodstream of patients receiving IV therapy. At low flow rates (≤ 1 ml/h) the alarm delay (time to an alert to the user) can be up to 2 h using conventional pressure threshold algorithms. In order to reduce alarm delays we developed and evaluated the performance of two new real-time occlusion detection algorithms and one co-occlusion detector that determine… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
5
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
4
1

Relationship

0
5

Authors

Journals

citations
Cited by 6 publications
(7 citation statements)
references
References 18 publications
0
5
0
Order By: Relevance
“…Occlusion is a blockage or closing of the infusion flow passage, usually referring to an interruption or impediment of an infusion tube due to a blockage, momentary closure, or obstruction of the passageway or blood vessel. Occlusion in the fluid tube of infusion pumps can prevent vital and time-critical medication from being delivered into the bloodstream of patients, resulting in patient harm caused by cessation of therapy [ 40 41 ]. Therefore, it is very important to immediately alert the patient and medical care providers to an occlusion event.…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…Occlusion is a blockage or closing of the infusion flow passage, usually referring to an interruption or impediment of an infusion tube due to a blockage, momentary closure, or obstruction of the passageway or blood vessel. Occlusion in the fluid tube of infusion pumps can prevent vital and time-critical medication from being delivered into the bloodstream of patients, resulting in patient harm caused by cessation of therapy [ 40 41 ]. Therefore, it is very important to immediately alert the patient and medical care providers to an occlusion event.…”
Section: Discussionmentioning
confidence: 99%
“…The measured alarm delay times were consistent with other studies, which showed that the mean time to trigger the occlusion alarm was significantly longer at low infusion rates [ 42 ]. The alarm delays of PS-1000 meet the overall time delay criteria of the conventional alarm threshold of the occlusion pressure (300–800 mmHg) [ 41 ]. This is consistent with other reports [ 41 42 ].…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…At the same time, most devices are also equipped with IP door open alarms, infusion completion alarms, operation forgetting alarms, and low battery alarms, etc. Pressure sensing technology can be used to detect blockages in infusion tubes 37 . Most bubble alarms use complex ultrasound or optical detection methods, which can detect bubbles as small as 0.1 mL of volume 10 .…”
Section: Abnormal Infusion Monitoringmentioning
confidence: 99%
“…Another approach is a drip bag scale, which can deduce the flow rate due to the reduced weight of the infusion bag [27]. Additionally, there are pressure sensing techniques to detect occlusions in infusion tubing [28]. Intelligent infusion pumps and the methods presented can reduce the number of medication errors, but not completely eliminate them [24].…”
Section: Introductionmentioning
confidence: 99%