2013
DOI: 10.1161/circinterventions.112.000790
|View full text |Cite
|
Sign up to set email alerts
|

MRI-Induced Stent Dislodgment Soon After Left Main Coronary Artery Stenting

Abstract: e58S everal studies have shown the safety of MRI after coronary stenting; however, few of them included patients soon after stenting. In this report, we describe a case of MRIinduced stent dislodgment from left main coronary artery (LMCA) 2 weeks after stenting. Case ReportA 56-year-old woman underwent coronary angiography for recurrence of angina 12 months after coronary artery bypass graft surgery. Coronary artery bypass graft surgery had been performed with left internal mammary artery grafted to left anter… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
4
0

Year Published

2014
2014
2020
2020

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 8 publications
(4 citation statements)
references
References 3 publications
0
4
0
Order By: Relevance
“…56 -58 To our knowledge, there has only been one unusual case report from Ontario, where a drug-eluting coronary stent (Cypher; Cordis Corp) in the left main coronary artery was dislodged when a 56-year-old patient underwent an MRI of the head in a 1.5 T scanner 2 weeks after stent implantation. 59…”
Section: Magnetic Resonance Imaging and Other Cardiac Devicesmentioning
confidence: 99%
See 1 more Smart Citation
“…56 -58 To our knowledge, there has only been one unusual case report from Ontario, where a drug-eluting coronary stent (Cypher; Cordis Corp) in the left main coronary artery was dislodged when a 56-year-old patient underwent an MRI of the head in a 1.5 T scanner 2 weeks after stent implantation. 59…”
Section: Magnetic Resonance Imaging and Other Cardiac Devicesmentioning
confidence: 99%
“…[56][57][58] To our knowledge, there has only been one unusual case report from Ontario, where a drug-eluting coronary stent (Cypher; Cordis Corp) in the left main coronary artery was dislodged when a 56-year-old patient underwent an MRI of the head in a 1.5 T scanner 2 weeks after stent implantation. 59 Bioprosthetic and mechanical valves, as well as annuloplasty rings, are labeled MR safe or MR conditional at 3 T or less, including the Starr-Edwards ball-and-cage mechanical heart valve that was once an absolute contraindication to MRI, but now considered MR conditional at 1.5 T. 60,61 Percutaneous transcatheter aortic valves are considered MR conditional at 1.5 and 3 T, when limiting SAR to <2 W/kg and imaging time to <15 minutes per pulse sequence. It has been shown that 3 T produces smaller forces on mechanical valves than a beating heart.…”
Section: Magnetic Resonance Imaging and Other Cardiac Devicesmentioning
confidence: 99%
“…The interaction may result in induced currents, device heating, and vibration of the device. Summing up, the different interactions between MRI and metallic implants are so strong that even MRI-induced stent dislodgment may be possible [10].…”
Section: Introductionmentioning
confidence: 99%
“…It is important to dispute that MRI caused coronary stent dislodgement in the patient described by Parthasarathy and colleagues 1 . The BX Velocity stent used in the Cypher device in their report is composed of 316L stainless steel 2 , which is weakly ferromagnetic.…”
mentioning
confidence: 98%