Mouse MRI using phased‐array coils

Abstract: The advantages of array coil imaging in human whole-body MR systems are well known and have gained relevance in many applications and MR techniques. In the field of small-animal studies, this concept has become increasingly important. In this work, three different phased-array coils for performing MRI on mice are presented. For imaging at 300 MHz, a four-channel receive-only phased-array coil is introduced. One two-channel and a four-channel transmit/receive setup operating at 750 MHz show the feasibility of a… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
36
0

Year Published

2008
2008
2019
2019

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 33 publications
(36 citation statements)
references
References 14 publications
(12 reference statements)
0
36
0
Order By: Relevance
“…A 3D FLASH sequence was used with the following parameters: 15 ms TR, 6 ms TE and 50 kHz receiver bandwidth. A total of 128 partitions (32.5 mm thick) were acquired with a FOV of 6.5 x 6.5 cm 2 and an acquisition matrix size of 256 x 256 in 8min12sec (NA=1). Fig.…”
Section: )mentioning
confidence: 99%
See 3 more Smart Citations
“…A 3D FLASH sequence was used with the following parameters: 15 ms TR, 6 ms TE and 50 kHz receiver bandwidth. A total of 128 partitions (32.5 mm thick) were acquired with a FOV of 6.5 x 6.5 cm 2 and an acquisition matrix size of 256 x 256 in 8min12sec (NA=1). Fig.…”
Section: )mentioning
confidence: 99%
“…Anti-aliasing was used in both slice and phase encoding direction. A total of 256 partitions (98 µm thick) were acquired using a double slab volume selection, one for each knee, with a FOV of 1.25 x 1.88 cm 2 and an acquisition matrix size of 256 x 384. Due to the simultaneous signal acquisition on four channels, the amount of data exceeded 4GBytes.…”
Section: )mentioning
confidence: 99%
See 2 more Smart Citations
“…For mouse imaging studies, the availability of arrays is mostly limited to linear array designs (5). The availability of mouse volume arrays is very limited when compared with the volume arrays for rat studies.…”
Section: Introductionmentioning