Abstract:Purpose:
To validate dose calculation and delivery accuracy of a recently introduced mono‐isocentric technique for the treatment of multiple brain metastases in a realistic clinical case.
Methods:
Anonymized CT scans of a patient were used to model a hollow phantom that duplicates anatomy of the skull. A 3D printer was used to construct the phantom of a radiologically bone‐equivalent material. The hollow phantom was subsequently filled with a polymer gel 3D dosimeter which also acted as a water‐equivalent mate… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.