1983
DOI: 10.1002/lsm.1900030107
|View full text |Cite
|
Sign up to set email alerts
|

Microbial Contamination by a Medical Carbon Dioxide Laser

Abstract: One of the advantages in using the carbon dioxide laser in medicine is the sterilization of the wound at the site of surgical intervention. In microbial studies, using the Sharplan Model 743 Medical Laser, we found substantial contamination of the area directly below the probe by viable bacteria and fungi. The levels of contamination varied from experiment to experiment, but were always substantial. The contamination is likely due to the stream of nitrogen gas emitted during and following laser irradiation in … Show more