Abstract:This paper presents the major impacts of the work developed in EMPIR Project -MeDD II, Metrology for Drug Delivery in the reference standards used for flow rate determination of drug delivery devices and the results obtained in each work package, with a special focus on the primary standards developed by the partners using different technologies. These new primary standards were validated via an inter-laboratory comparison using two different type of flow meters and a precision syringe pump, in a range from 15… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.