Metrologically traceable quantification of trifluoroacetic acid content in peptide reference materials by19F solid-state NMR

Abstract: Although solution-state NMR is frequently used in metrologically-traceable quantification studies, this is not the case for solid-state NMR. However, solid-state NMR allows quantification of substances without the need of dissolution, providing a truly non-destructive approach, and extending metrologically-traceable quantitative NMR to sample classes that are difficult to characterize in solution. In this contribution we present a thorough and rigorous protocol for 19 F quantitative solid-state NMR employing a… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0

Year Published

2019
2019
2022
2022

Publication Types

Select...
5

Relationship

2
3

Authors

Journals

citations
Cited by 5 publications
(2 citation statements)
references
References 80 publications
0
2
0
Order By: Relevance
“…The probe was tuned to the 19 F Larmor frequency given by 188.7 MHz prior to every measurement. The published protocol for 19 F quantitative solid-state NMR 26 was employed, specically metrological traceability was achieved by using 3,5bis(triuoromethyl)benzoic acid (BTFMBA) as external calibrant. Furthermore, we employed the ERETIC (Electronic REference To access In vivo Concentrations) and EASY (Elimination of Artifacts in NMR Spectroscopy) methods.…”
Section: -Nitrobenzaldehyde Assaymentioning
confidence: 99%
“…The probe was tuned to the 19 F Larmor frequency given by 188.7 MHz prior to every measurement. The published protocol for 19 F quantitative solid-state NMR 26 was employed, specically metrological traceability was achieved by using 3,5bis(triuoromethyl)benzoic acid (BTFMBA) as external calibrant. Furthermore, we employed the ERETIC (Electronic REference To access In vivo Concentrations) and EASY (Elimination of Artifacts in NMR Spectroscopy) methods.…”
Section: -Nitrobenzaldehyde Assaymentioning
confidence: 99%
“…The assigned values can be confirmed by microanalysis (% C, H, N content) or by TGA for total non-volatile inorganic impurities. Alternatively, accurate TFA quantification can be achieved by liquid phase quanti tative 19 F-NMR [82,100] or solid state quantitative 19 F-NMR [101].…”
Section: (C)residual Organic Solvent (Volatile Organic Compounds)mentioning