2019
DOI: 10.1016/j.fuel.2019.01.176
|View full text |Cite
|
Sign up to set email alerts
|

Metals determination in crude oil by inductively coupled plasma optical emission spectrometry using nanoemulsification as sample preparation

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

2
3
0

Year Published

2020
2020
2024
2024

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 18 publications
(5 citation statements)
references
References 41 publications
2
3
0
Order By: Relevance
“…It is important to note that the limit of detection (LOD) was chosen as one of the figures used to assess the merits of the ILA-EIEB, since it is an analytical parameter that is closely related to sensitivity. The proposed ILA-EIEB method showed 2.68, 0.03, 8.74 and 1.399 μg/L LODs for Ba, Na, Ni and V, respctively, which were quite comparable with literature reported studies [ [10] , [32] , [37] , [43] , [44] , [45] ]. Furthermore, the newly developed ILA-EIEB method was accurate (80-101) and precise (1.9-4.7) for all the invesigated metals (Ba, Na, Ni and V).…”
Section: Resultssupporting
confidence: 88%
See 2 more Smart Citations
“…It is important to note that the limit of detection (LOD) was chosen as one of the figures used to assess the merits of the ILA-EIEB, since it is an analytical parameter that is closely related to sensitivity. The proposed ILA-EIEB method showed 2.68, 0.03, 8.74 and 1.399 μg/L LODs for Ba, Na, Ni and V, respctively, which were quite comparable with literature reported studies [ [10] , [32] , [37] , [43] , [44] , [45] ]. Furthermore, the newly developed ILA-EIEB method was accurate (80-101) and precise (1.9-4.7) for all the invesigated metals (Ba, Na, Ni and V).…”
Section: Resultssupporting
confidence: 88%
“…Furthermore, the newly developed ILA-EIEB method was accurate (80-101) and precise (1.9-4.7) for all the invesigated metals (Ba, Na, Ni and V). It is worthy to indicate that, several studies have reported on EIEB, where high volumes of carcinogenic organic solvents (xylene and hexane) and concentrated corrosive acids (HNO 3 and HCl) were used, making the extraction environmentally unfriendly [ 35 , [43] , [46] , [47] ]. It was also noted that, the extraction of V was challenging, and these shortcomings were reported in literature, where extraction effeciencies of 50–74% were obtained [ 10 ].…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…The most optimum digestion conditions were predicted to be 0.1 g sample mass, 245 °C digestion time, 25 min digestion time and 5 M of H 2 O 2 based on the quadratic equations and the surface plots obtained from the Minitab 18 software. The response optimiser was also used to predict the optimum conditions of microwave digestion and the findings were in agreement with quadratic equations and the surface plots results 31 . The response surface obtained from Minitab 2018 conditions indicated good accuracy of 106.8, 111.7, 90.0 and 109.8% for Ba, Na, Ni and V at 95% confidence level with precision of 2.9, 0.8, 2.7 and 4.1 for Ba, Na, Ni and V, respectively (results A).…”
Section: Resultssupporting
confidence: 59%
“…Both rural and urban inhabitants in Africa lack access to safe drinking water. The majority of turbidity-related issues will be alleviated by removing colloidal and suspended particles from the water (Vieira Luiza, et al, 2019). The majority of particles are too tiny to settle by weight and are able to pass through the pores of the majority of lters.…”
Section: Introductionmentioning
confidence: 99%